The First and Only
Full-Rescue AED

Zoll AEDPLUS

 

Features

Specifications

Comparison

Accessories


Specifications
Defibrillator
Waveform Rectilinear Biphasic
Defibrillator Charge Hold Time 30 seconds
Energy Selection Automatic preprogrammed selection
Patient Safety All patient connections are electrically isolated
Charge Time Less than 10 seconds with new batteries
Electrodes ZOLL CPR-D padz or stat padz II
CPR: Metronome Rate Variable 60 to 100 CPM
Depth 1/2" to 3"
Defibrillator Protected ECG Circuitry
Display Format Optional LCD with Moving Bar
Size 2.6" x 1.3"
Viewing Time 2.6 seconds
Display Sweep Speed 25 mm/sec; 1"/sec
Battery Capacity Typical new (20°C) = 5 years (300 shocks) or 1.5 hours continuous Monitoring/Defibrillation.
PC Min. Requirements: Windows® 98, Windows® 2000, Windows®NT, Windows® XP, IBM-compatible PII with 16550 UART (or higher) computer. 64MB RAM. VGA monitor or better. CD-ROM drive. IrDA port. 2MB disk space.
Defibrillation Advisory: Evaluates electrode connection and patient ECG to determine if defibrillation is required. Shockable Rhythms: Ventricular fibrillation with average amplitude >100 microvolts and wide complex ventricular tachycardia with rates greater than 150 BPM. For ECG Analysis Algorithm Accuracy for sensitivity and specificity performance, refer to AED Plus Administrator's Guide.

Device
Size (H x W x D) 5.25"x 9.50" x 11.50"
Weight 6.7 lbs.
Power User Replaceable Batteries. 10 -Type 123A Photo Flash lithium manganese dioxide batteries.
Device Classification Class II and internally powered per EN60601-1
Design Standards Meets applicable requirements of UL 2601, AAMI DF-39, IEC 601-2-4, EN60601-1, IEC60601-1-2.

Environment
Operating Temperature PS Model: 32. to 122.F; 0. to 50.C
PA Model: 50. to 104.F; 10. to 40.C
Storage
Temperature
PS Model: -22. to 158.F; -30. to 70.C
PA Model: 32. to 122.F; 0. to 50.C
Humidity 10 to 95% relative humidity, non-condensing
Vibration MIL Std. 810F, Min. Helicopter Test
Shock PS Model: IEC 68-2-27; 100G
PA Model: IEC 68-2-27; 50G
Altitude PS Model: -300 to 15,000 ft.; -91m to 4573m;
PA Model: -300 to 7,500 ft.; -91m to 2287m
Particle and Water Ingres IP-55

CPR-D Padz
Shelf Life 4 Years
Conductive Gel Polymer Hydrogel
Conductive Element Tin
Packaging Multilayer foil laminate pouch
Impedance Class Low
Cable Length 36 in (91 cm)
Sternum: Length: 6.1 in (15.5 cm); Width: 5.0 in (12.7 cm); Length, conductive gel: 3.5 in (8.9 cm); Width, conductive gel: 3.5 in (8.9 cm); Area, conductive gel: 12.3 sq in (79.0 sq cm)
Apex: Length: 6.1 in (15.5 cm); Width: 5.6 in (14.1 cm); Length, conductive gel: 3.5 in (8.9 cm); Width, conductive gel: 3.5 in (8.9 cm); Area, conductive gel: 12.3 sq in (79.0 sq cm)
Complete Assembly Folded Length: 7.6 in (19.4 cm); Folded width: 7.0 in (17.8 cm); Folded height: 1.5 in (3.8 cm)
Design Standards Meets applicable requirements of ANSI/AAMI/ISO DF-39-1993.


About Us Products Contact Us FAQ Links Home Home About Us Products Contact Us FAQ Links