| Physical |
| Size |
8" x 7.25" x 1.9" (20cm x 18.4cm x
4.8cm) |
| Weight |
2.4 lbs. (1.1kg) including Pad-PAK™ Battery
|
| Defibrillator |
| Waveform |
SCOPE® (self-compensating output pulse envelope)
Biphasic Escalating waveform. Optimized biphasic waveform
compensates energy, slope & envelope for patient
impedance. |
| Energy Selection |
Pre-configured automated protocol (escalating from
100J, 150J, to 200J). |
| Charge Time |
Less than 7 seconds (100J) and less than 12 seconds
(200J) from a SA message with new Pad-Pak™. |
| Shock Cycle Time |
<60 seconds for three analyze/shock cycles. |
| Controls |
On/Off and Shock Button |
| Patient Analysis System |
| Method |
Evaluates patient�s ECG, signal quality, electrode
contact integrity and patient impedance to determine
if defibrillation is required. |
| Sensitivity/Specifically |
Meets AAMI DF 80 |
| Display |
| Visual Prompts |
Attach Pads, Stand Clear, Perform CPR, Shock Now,
Self Tests Pass - Ready State. |
| Indicators |
Low battery, (at least 10 discharges remaining), audible
prompt (alerts user of electrode disconnect), service
icon stops flashing if self-tests fail, no battery,
or when service required. |
| Voice and Icon
Prompts |
Extensive voice and icon prompts guide the user through
the operation sequence. |
| Languages |
English |
| Environmental |
| Operating Temperature |
0° to 50° C (+32° to +112° F) |
| Storage Temperature |
-10° to 50° C |
| Relative Humidity |
5% to 95% (non-condensing) |
| Water Resistance |
IEC 60529/EN 60529 IP44 |
| Altitude |
0 to 15,000 feet (0 - 4,575 meters) |
| Shock |
MIL STD 810F Method 516.5, procedure I (40G's) |
| Vibration |
MIL STD 810F Method 514.5
Category 4 Truck Transportation - US Highways
Category 7 Aircraft - Jet (737 & General Aviation Exposure) |
| EMC |
| EN 60601-1-2, Second Edition: 2001 |
| Radiated Emissions |
CISPR11: 1997 +1A:1999+A2:2001 Group 1 Class B |
| Electrostatic Discharge Immunity |
EN61000-4-2:1995 (8KV)+A2:2000
|
| RF Immunity |
EN61000-4-3:1996. 80 MHz-2.5GHz (10V/m)+A1:1998+A2:2000 |
| Magnetic Field Immunity |
EN61000-4-8:1993 (3 A/m)+A1:2000 |
| Aircraft |
RTCA / DO - 160D: 1997, Section 21 (Category M) |
| Event Documentation |
| Type |
Internal Memory |
| Memory Capacity |
45 minutes of ECG (full disclosure) and event/incident
recording |
| Playback Capabilities |
Custom USB cable directly connected to PC and SAVER®
Windows-based data review software |
| Ped-PAK™ Battery |
| Shelf Life |
3 years |
| Weight |
0.44 lbs. (0.2kg) |
| Size |
3.93" x 5.24" x .94"
(100mm x 133mm x 24mm) |
| Battery Type |
Lithium Manganese Dioxide
18V, 0.8 Amp Hrs |
| Capacity |
>30 shocks at 200J or 6 hours of continuous monitoring. |
| Standby Life |
3 years |
| Electrodes |
Samaritan® disposable defibrillation electrodes
are supplied standard with each Pad-PAK™ |
| Active Gel Area |
100cm² |
| Cable Length |
3.5 ft (1m) |
| Electrode Shelf Life |
3 years |